510(k) Data Aggregation

    K Number
    DEN200015

    Validate with FDA (Live)

    Date Cleared
    2020-07-30

    (150 days)

    Product Code
    Regulation Number
    866.3183
    Type
    Direct
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    cobas EBV is an in vitro nucleic acid amplification test for the quantitation of Epstein-Barr virus (EBV) DNA in human EDTA plasma on the cobas 6800/8800 Systems.

    cobas EBV is intended for use as an aid in the management of EBV in transplant patients. In patients undergoing monitoring of EBV, serial DNA measurements can be used to indicate the need for potential treatment changes and to assess response to treatment.

    The results from cobas EBV are intended to be read and analyzed by a qualified licensed healthcare professional in conjunction with clinical signs and symptoms and relevant laboratory findings. Negative test results do not preclude EBV infection or EBV disease. Test results must not be the sole basis for patient management decisions.

    cobas EBV is not intended for use as a screening test for donors of blood or blood products or human cells, tissues, and cellular and tissue-based products (HCT/Ps).

    Device Description

    cobas EBV is a quantitative test performed on the cobas 6800 System and cobas 8800 System. cobas EBV enables the detection of EBV DNA in plasma specimens. The cobas EBV assay is a dual target assay, with both targets using the same dye. The DNA Internal Control, used to monitor the entire sample preparation and PCR amplification process, is introduced into each specimen during sample processing. cobas EBV enables the detection and quantitation of EBV DNA in EDTA plasma from solid organ transplant patients (SOT) and from hematopoietic stem cell transplant (HSCT) patients. The test is intended for use as an aid in the management of SOT patients and HSCT patients.

    The cobas EBV consists of:

    • Proteinase Solution ●
    • DNA Quantitation Standard (DNA QS) ●
    • Elution Buffer ●
    • Master Mix Reagent 1
    • . EBV Master Mix Reagent 2

    The EBV viral load is quantified against a non-EBV DNA quantitation standard (DNA-OS), which is introduced into each specimen during sample preparation. The DNA-QS also functions as an internal control for sample preparation and the PCR amplification process.

    In addition, the test utilizes the following separately packed and sold control materials:

      1. cobas EBV Positive Control Kit:
      • . EBV Low Positive Control (EBV L(+)C)
      • EBV High Positive Control (EBV H(+)C) ●

    The positive control contains phage packaged EBV DNA in normal human plasma and serves as a control for the cobas EBV test.

      1. cobas Negative Control Kit:
      • cobas Buffer Negative Control (BUF (-) C) ●

    Testing with the cobas EBV test requires the following materials that are not provided:

    • cobas OMNI Reagents: Including the following reagents used for specimen ● processing, PCR and detection:
    • cobas EBV Assay Specific Analysis Package (ASAP) software .

    The cobas EBV test uses sample preparation (nucleic acid extraction and purification) followed by PCR amplification and detection, all steps are fully automated by the cobas 6800/8800 platform.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the studies that prove the device meets these criteria, based on the provided text:

    Acceptance Criteria and Device Performance

    CriteriaReported Device Performance
    PrecisionStandard Deviation (log10 transformed titer): Ranged from 0.02 to 0.17 across various concentrations (Table 1). Lognormal Percent Coefficient of Variation (%CV): Ranged from 7% to 43% for positive panel members (Table 2).
    ReproducibilityTotal Lognormal Coefficient of Variation (% CV): Ranged from 13.7% to 46.16% among positive panel members when tested across three reagent lots, three test sites, and three instruments (Table 3).
    Linearity (Genotype 1)Linear Range: 35 IU/mL (LLoQ) to 1.0E+08 IU/mL (ULoQ). Accuracy: Within ± 0.15 log10 IU/mL (Mean Square Error) across the linear range. Best-fitting model: 1st order for individual lots/panel types, 2nd order for combined data (minor difference to 1st order: ≤ ± 0.01 log10 IU/mL). (Table 4, Figure 1)
    Linearity (Genotype 2)Accuracy: Within ± 0.12 log10 IU/mL (Mean Square Error) across the linear range, with minor differences (–0.06 log IU/mL to +0.08 log IU/mL) between 1st and 3rd order regression models. (Figure 2)
    TraceabilityQuantitation values for the cobas EBV calibration panel and RMS EBV Secondary Standard showed deviations not more than 0.15 log10 IU/mL from expected values, demonstrating traceability to the 1st WHO International Standard for EBV (NIBSC 09/260). (Figure 3) Regression equations: EBV Calibration Panel: y = 1.000x - 0.002; R2 = 1.000. EBV 1st WHO Standard: y = 0.975x + 0.159; R2 = 0.983.
    Clinical Specimen StabilityWhole blood (EDTA-plasma tubes) stable for up to 24 hours at 2-25°C. Plasma stable for 24 hours at 2-30°C, then up to 6 days at 2-8°C, or up to 6 months at -15 to -80°C. Plasma stable for up to four freeze/thaw cycles. Mean log10 titers at different time points/conditions were within ±0.5 log10 of the reference (T0).
    Open Kit and On-board StabilityOpen test-specific reagent cassettes stable for up to 90 days at 2-8℃ (Open Kit) and up to 40 hours at 37°C (On-Board). Reusable for up to 40 runs.
    Reagent StabilityShelf-life stability claim of 12 months when stored at 2-8°C (supported by real-time stability data).
    Limit of Detection (LoD)WHO EBV Standard (Genotype 1): LoD = 18.8 IU/mL by Probit analysis (highest LoD across lots). LoD by 95% hit rate was (b) (4) for all lots. (Table 5) EBV Genotype 2: LoD verified at 18.8 IU/mL with a hit rate of 95% or higher. (Table 6) Plasma vs Buffer (GSD): Comparable LoD performance in Plasma and GSD (Hit rate 98.4% at 1.5xLoD, 92.1% at 1.0xLoD). (Table 7)
    Lower Limit of Quantitation (LLoQ)LLoQ: 35 IU/mL. Determined based on Total Analytical Error (TAE) and difference between two measurements. (Table 8)
    Cross-reactivityEBV-negative samples: Negativity rate was high (not explicitly stated as 100% but implied by lack of reported positives). EBV-positive samples: Mean log10 titer of positive EBV samples with 35 potential cross-reacting microorganisms was within ±0.5 log10 of the positive spike control. (Table 9)
    Endogenous InterferenceEBV-negative samples: All produced valid negative results. EBV-positive samples: Mean log10 titer of positive EBV samples with interfering substances (NaOH2, Albumin, Bilirubin, Human DNA, Hemoglobin, Triglycerides) was within ±0.05 log10 of the spike control. (Table 10)
    Exogenous InterferenceEBV-negative samples: 100% negativity rate. EBV-positive samples: Mean log10 titer of positive EBV samples with 24 common drugs was observed to be within acceptable range (implied by table values, e.g., Max Mean Difference in log10 Titer of -0.03 for Ethanol SC). (Table 11)
    Cross-Contamination0% cross-contamination rate (upper one-sided 95% CI: 1.24%) from testing 240 replicates of EBV-negative matrix sample in checkerboard configuration with high-titer samples.
    Concordance with Comparator EBV TestOverall Column Percent Agreement: Ranged from 82.5% to 100% depending on analyte concentration (for valid samples on both assays, n=464). (Table 12) Negative Percent Agreement (NPA): 95.4% (95% CI: 84.2%-99.4%) with comparator EBV negative samples (n=43). Agreement at clinical thresholds: High agreement, e.g., 98.0% (<LLOQ), 100.0% (<3.0 Log10 IU/mL), 100.0% (<4.0 Log10 IU/mL). (Table 13) Deming Regression: Y = -0.45 + 1.025X, R2 = 0.91, indicating a strong correlation with a systematic difference within acceptable range (intercept 95% CI: -0.456 to -0.104, within ±0.6 log10 IU/mL). (Figure 4)

    Study Details

    2. Sample Size and Data Provenance

    • Test Set (Clinical Study):
      • Sample Size: 464 samples (439 neat or diluted clinical samples from 72 transplant subjects, and 25 contrived samples).
      • Data Provenance: Clinical samples from a major transplant center in the United States (implied prospective or retrospective collection for comparison, but not explicitly stated as either for the clinical samples themselves). Contrived samples were prepared in the lab.

    3. Number of Experts and Qualifications (for Ground Truth)

    • The document does not specify the number of experts or their qualifications for establishing the ground truth for the clinical test set. The comparator EBV test is described as a "validated well-established comparator nucleic acid test currently used in clinical practice at a major transplant center in the United States." This implies expert-driven clinical practice but no explicit panel for ground truth review is mentioned.

    4. Adjudication Method (for Test Set)

    • None specified for the concordance study. The primary method was direct comparison to a "validated well-established comparator nucleic acid test." Discordant results were defined and analyzed based on a quantitative difference (more than one box away from the diagonal in the concordance table). DNA sequencing was used on representative samples for sequence mismatches in large discrepancies, which is a form of expert adjudication for cause but not for initial ground truth determination.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, a MRMC comparative effectiveness study was not done. The study focused on the analytical and clinical concordance of the automated device with an existing laboratory-developed test (LDT), not on human reader performance with and without AI assistance.

    6. Standalone Performance Study (Algorithm Only)

    • Yes, a standalone study was done. The entire analytical performance section (Precision, Reproducibility, Linearity, Traceability, Stability, LoD, LLoQ, Analytical Specificity, Cross-Contamination) evaluates the cobas EBV test (algorithm and instrument) in a standalone capacity, without human intervention in the result generation. The clinical concordance study also compares the device's output to a comparator, effectively assessing its standalone performance.

    7. Type of Ground Truth Used

    • For Analytical Studies: Spiked biological samples (e.g., EBV phagemid, EBV cell culture supernatant, WHO EBV Standard) with known concentrations, or EBV-negative plasma samples, were used to establish ground truth for parameters like LoD, Linearity, and Specificity.
    • For Clinical Concordance Study: The ground truth was established by a "validated well-established comparator nucleic acid test" currently used in clinical practice at a major transplant center. This is a form of reference method/comparator assay ground truth.

    8. Sample Size for Training Set

    • The document does not explicitly state the sample size used specifically for the training set. It describes the reagents and the test principle, but not the specific data used for initial algorithm development or 'training' in the machine learning sense. The performance studies use various panels and clinical samples for validation of the finalized assay.

    9. How Ground Truth for Training Set Was Established

    • As the document does not specify a distinct 'training set' or 'algorithm training', it also does not explain how ground truth was established for it. For a PCR-based quantitative assay, the 'training' often involves optimizing primer/probe design and assay conditions against well-characterized reference materials (like the WHO standard or cultured virus) to achieve desired analytical performance parameters before formal validation studies.